• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2020
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be evaluated by syncardia.The results of the evaluation will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver did not pass the system check procedure.
 
Manufacturer Narrative
The reported issue was confirmed by review of the patient data file which revealed a system check test failure for the 9-volt battery reading below acceptable limits (6.500 v - 10.000 v).The 9-volt battery is replaced at every service (90-days of use interval or 2-year calendar interval).The battery that was replaced at the most recent service interval had an expiration date of march 2021.At the time that the 9v battery failure occurred, the battery was not expired.It is suspected that the 9v battery experienced an unexpected cell depletion malfunction.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
sagar pimpalwar
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key10503234
MDR Text Key207528945
Report Number3003761017-2020-00192
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-