Model Number 397002-001 |
Device Problems
Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/26/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be evaluated by syncardia.The results of the evaluation will be provided in a follow-up mdr.(b)(4) initial.
|
|
Event Description
|
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver did not pass the system check procedure.
|
|
Manufacturer Narrative
|
The reported issue was confirmed by review of the patient data file which revealed a system check test failure for the 9-volt battery reading below acceptable limits (6.500 v - 10.000 v).The 9-volt battery is replaced at every service (90-days of use interval or 2-year calendar interval).The battery that was replaced at the most recent service interval had an expiration date of march 2021.At the time that the 9v battery failure occurred, the battery was not expired.It is suspected that the 9v battery experienced an unexpected cell depletion malfunction.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
|
|