• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC PRECISION QID; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC PRECISION QID; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2020
Event Type  malfunction  
Manufacturer Narrative
Upon extended investigation, it was determined that the serial number provided by the customer ((b)(4)) was not a valid serial number.An extended investigation has been conducted, which involved a manufacturing review (including 5 years of device history records (dhrs)), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation did not identify any indication that the product did not meet specification.If the product is returned, an investigation will be performed.Note: the device manufacturer date for the reported meter serial number is unknown.The date entered is the date abbott diabetes care became aware of the event.
 
Event Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION QID
Type of Device
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
tara williamson
1360 south loop road
alameda, CA 94502-7001
5108644472
MDR Report Key10514906
MDR Text Key207291694
Report Number2954323-2020-06129
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
PMA/PMN Number
K971812
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Date Manufacturer Received09/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-