Model Number 0450000000
Device Problems Suction Problem (2170); Insufficient Information (3190)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/14/2020
Event Type  Injury  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that the procedure was cancelled and there is a need for a follow up procedure.
 
Manufacturer Narrative
Alleged failure: (b)(4), rep, was not suctioning during the case the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be 1) intermittent software error.2) shaver could be clogged.3) joint level set up / set up.4) clamp closed on tubing.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Event Description
It was reported that the procedure was cancelled and there is a need for a follow up procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, CROSSFLOW CONSOLE
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10515678
MDR Text Key206362547
Report Number0002936485-2020-00369
Device Sequence Number1
Product Code HRX
UDI-Device Identifier07613327058048
UDI-Public07613327058048
Combination Product (y/n)N
PMA/PMN Number
K123441
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0450000000
Device Catalogue Number0450000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Date Manufacturer Received08/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-