Model Number 0450000000 |
Device Problems
Suction Problem (2170); Insufficient Information (3190)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/14/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
|
Event Description
|
It was reported that the procedure was cancelled and there is a need for a follow up procedure.
|
|
Manufacturer Narrative
|
Alleged failure: (b)(4), rep, was not suctioning during the case the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be 1) intermittent software error.2) shaver could be clogged.3) joint level set up / set up.4) clamp closed on tubing.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
|
Event Description
|
It was reported that the procedure was cancelled and there is a need for a follow up procedure.
|
|