Model Number 195215 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/20/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) replaced the level sensor.The unit operated to the manufacturer's specifications.The suspect device was returned to the manufacturer for further evaluation.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the level sensor had a deteriorating membrane and was intermittently working.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|