D10: date returned to manufacturer is 9/16/2020.H10: one (1) used partial list # 011-0j988-01, transpac¿ trifurcated monitoring kit, 60" tubing, 3 3 ml flush device, microdrip; lot # 3964739 was received for evaluation.The reported complaint of disconnection was confirmed on the returned set.During visual inspection, the 60" pressure tubing was received broken from the winged male luer.The winged male luer was not received for evaluation.It is unknown how and when the tubing breakage had occurred.The packaging was observed and no indentation was observed which could lead to the to the reported complaint.The returned trifurcated monitoring set was primed and pressure leak tested and no leaks were observed.The probable cause for the broken pressure tubing is unknown.
|