• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION KIT Back to Search Results
Model Number IPN046311
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/25/2020
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the plunger is air leaking.The doctor(anesthetist) faced 2 incidents during the night of (b)(6) 2020.With 2 different lot numbers.Lot#:71f20e0804 and 71f20e0956.There was no consequence for the two female patients but a risk.During the week 35 (17 to 23 aug 20),another doctor(anesthetist) faced 2 incidents with lot# 71f20e0804 and 71f20e095.The consequences for the 2 female patients were: dura mater breaches so a blood patch had to be applied for each.
 
Event Description
It was reported that the plunger is air leaking.The doctor(anesthetist)faced 2 incidents during the night of wednesday (b)(6) 2020.With 2 different lot numbers.Lot#:71f20e0804 and 71f20e0956.There was no consequence for the two female patients but a risk.During the week 35 (17 to 23 aug 20),another doctor(anesthetist) faced 2 incidents with lot# 71f20e0804 and 71f20e095.The consequences for the 2 female patients were: dura mater breaches so a blood patch had to be applied for each.
 
Manufacturer Narrative
Qn# (b)(4).A device history record review was performed on the lor syringe with no relevant findings.A design history review was performed for part # kz-05501-002 as a part of this complaint investigation.Per eco-057562 (released 12-mar-2020), supplier (preox) made the following changes: optional component materials/mold locations: option 1: barrel: polypropylene - profax pf-535 lyondell-basell plunger: polypropylene 100-ga12 ineos olefins & polymers (molded at fleima plastic) blue stopper: silicone rubber (molded at et elastomer technik).Option 2: barrel: polypropylene - profax pf-535 lyondell-basell plunger: polypropylene profax pf-531 lyondell-basell (molded at gpe) blue stopper: silicone rubber (molded at psilkon).Lubricant: - the i.D.Of the barrel lubricated w/ medical grade silicone (polydimethylsiloxan) and amount will not exceed 0.25mg per sq centimeter.These effective changes did impact product design and material.A corrective action is not required at this time as the root cause for this complaint investigation could not be determined based upon the information provided and without the actual sample.Complaint verification testing could not be performed as no sample was provided for analysis.A device history record review was performed on the lor syringe with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without the sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Type of Device
ANESTHESIA CONDUCTION KIT
MDR Report Key10536599
MDR Text Key207340984
Report Number3006425876-2020-00787
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/11/2022
Device Model NumberIPN046311
Device Catalogue NumberJC-05400-B
Device Lot Number71F20E0956
Was Device Available for Evaluation? No
Date Manufacturer Received10/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-