• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 72204064
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Unintended Radiation Exposure (4565)
Event Date 08/20/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during an operative hysteroscopy, inside the uterus, the device was likely torqued which, therefore caused the blade to significantly bend and then it broke off.The internal cutting blade broke off inside the patient's uterus.An x-ray was done.All fragments were successfully retrieved and the procedure was ended.
 
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the unit had an inner blade fracture.It was reported that the device had a component that disengaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
MDR Report Key10538374
MDR Text Key207096567
Report Number1282497-2020-00036
Device Sequence Number1
Product Code HIH
UDI-Device Identifier10884521744097
UDI-Public10884521744097
Combination Product (y/n)N
PMA/PMN Number
K161763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/05/2021
Device Model Number72204064
Device Catalogue Number72204064
Device Lot Number5014606
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Date Manufacturer Received03/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age45 YR
-
-