• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC INFINITY OCCIPITOCERVICAL UPPER THORACIC SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC INFINITY OCCIPITOCERVICAL UPPER THORACIC SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 3602536
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/12/2020
Event Type  malfunction  
Manufacturer Narrative
Product analysis#: 703805582:3602536, lot# rm18a013.Visual and optical examination identified that the hex of the cone set screw has been stripped.The stripped cone set screw is consistent with over-torque.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider via a manufacturer representative regarding a patient with trauma.It was reported that the lateral set screw stripped as surgeon was attempting to final tighten.The device was replaced, and will be returned.There were no symptoms reported.There were no further complications reported regarding the event.On (b)(6) 2020, (rep) additional information was received from manufacturer representative regarding a patient with occiput to c4 p osterior fusion (levels implanted-occiput to c4) due to trauma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY OCCIPITOCERVICAL UPPER THORACIC SYSTEM
Type of Device
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
tricha miles
1800 pyramid place
memphis, TN 38132
7635140379
MDR Report Key10539976
MDR Text Key207161920
Report Number1030489-2020-01291
Device Sequence Number1
Product Code KWP
UDI-Device Identifier00643169527676
UDI-Public00643169527676
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163375
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3602536
Device Catalogue Number3602536
Device Lot NumberRM18A013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/11/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age10 YR
Patient Weight97
-
-