MAUDE Adverse Event Report: SMITH & NEPHEW, INC. AHTB ACTIVE HEEL TRACTION BOOT; APPARATUS, TRACTION, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. AHTB ACTIVE HEEL TRACTION BOOT; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Model Number 72202682
Device Problem Positioning Problem (3009)
Patient Problem No Patient Involvement (2645)
Event Date 08/24/2020
Event Type  malfunction  
Event Description
It was reported that before an arthroscopy, the foam sole of the active heel traction boot has come away from the back of the boat causing issues when securing the patients foot.The procedure was completed without delay using the same device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, which was used in a procedure, was not returned for evaluation.A relationship between the device and the reported incident could not be confirmed.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related failures.Factors that are known to contribute to the alleged fault/failure may be an adhesive/velcro failure.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.If the product is returned in the future the complaint can be reopened and evaluated.Our quality department will continue to monitor for trends.
 
Manufacturer Narrative
(b)(4).H3, h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection revealed the adhesive backing of the foam insert has detached from the shell and the shell was cracked on one side.The complaint was confirmed and the root cause was determined to be a component failure of the adhesive strip.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The complaint history review found further instances of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHTB ACTIVE HEEL TRACTION BOOT
Type of Device
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key10546078
MDR Text Key207308205
Report Number3003604053-2020-00107
Device Sequence Number1
Product Code HST
UDI-Device Identifier00885554024807
UDI-Public00885554024807
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup,Followup
Report Date 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202682
Device Catalogue Number72202682
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Date Manufacturer Received06/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-