Model Number N/A
Device Problems Separation Failure (2547); Failure to Cut (2587)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the cable of the sternal closure device would not cut and the plate was implanted without the cable.No adverse events have been reported as a result of the malfunction.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6.Visual examination of the product showed that the metal band was successfully locked in place by the thumb cam at the top of the tensioner.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERNALOCK 360 MULTI-IMPLANT SYSTEM
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key10553439
MDR Text Key207534488
Report Number0001032347-2020-00446
Device Sequence Number1
Product Code HRS
UDI-Device Identifier00841036190668
UDI-Public00841036190668
Combination Product (y/n)N
PMA/PMN Number
K151019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number74-0004
Device Lot Number066030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2020
Was the Report Sent to FDA? No
Date Manufacturer Received03/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Weight81
-
-