ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problems Therapeutic or Diagnostic Output Failure (3023); Intermittent Communication Failure (4038)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/16/2020
Event Type  malfunction  
Event Description
The recipient reportedly experienced loss of lock.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.In addition, broken antenna coil wires were observed.The photographic imaging inspection confirmed a broken antenna coil.System lock could not be obtained at any spacing.The no lock condition prevented some electrical tests from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.This device had broken antenna coil wires.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10560183
MDR Text Key207904221
Report Number3006556115-2020-00978
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 08/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/01/2012
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age11 YR
-
-