• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE Back to Search Results
Model Number DBEC-125
Device Problem Device Remains Activated (1525)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation conclusion: the oad was received at csi for analysis.Visual examination revealed shipping damage, but no other visible damage was observed.The oad operated as intended during functional testing.Additional analysis revealed moisture ingress within the start switch.The exact root cause of the fluid ingress could not be determined.If fluid ingress occurred during the procedure, then the start switch could have become unresponsive or experienced an unintended button press event.At the conclusion of the failure analysis investigation, the reported inability of the oad to stop spinning was confirmed, but the root cause could not be determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Event Description
During the third treatment with a diamondback coronary orbital atherectomy device (oad), the device would not stop spinning when the on/off button was pressed.The pump power cord was pulled out of the wall to turn off the oad.The procedure was completed with stent placement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK CORONARY ORBITAL ATHERECTOMY SYSTEM
Type of Device
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
morgan hill
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key10562632
MDR Text Key207810086
Report Number3004742232-2020-00282
Device Sequence Number1
Product Code MCX
UDI-Device Identifier10850000491356
UDI-Public(01)10850000491356(17)220430(10)324203
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model NumberDBEC-125
Device Catalogue Number7-10060-01
Device Lot Number324203
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
Patient Weight70
-
-