Device evaluation conclusion: the oad was received at csi for analysis.Visual examination revealed shipping damage, but no other visible damage was observed.The oad operated as intended during functional testing.Additional analysis revealed moisture ingress within the start switch.The exact root cause of the fluid ingress could not be determined.If fluid ingress occurred during the procedure, then the start switch could have become unresponsive or experienced an unintended button press event.At the conclusion of the failure analysis investigation, the reported inability of the oad to stop spinning was confirmed, but the root cause could not be determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|