The reported event was inconclusive.No sample was returned for evaluation.The potential root cause for this failure mode could be user related (example: contact with sharp object)/ mechanical failure/ operator error/ thin rubberize layer.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "method for use: (1) do not inflate the balloon in the urethra.(the urethra may be injured).(2) do not pull the catheter hard.(the bladder/urethra maybe injured).2.Applicable patients: (1) patients with delirium who might pull out catheter (when patient tugs at catheter unconsciously, the bladder and urethra may be damaged).Contraindications: 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (4) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients: (1) do not use on patients who are or have been allergic to natural rubber latex." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|