Catalog Number 0165L16
Device Problems Leak/Splash (1354); Decrease in Pressure (1490)
Patient Problem No Patient Involvement (2645)
Event Date 08/31/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that water leaked from the catheter during a pretest.
 
Event Description
It was reported that water leaked from the catheter during a pretest.
 
Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.The potential root cause for this failure mode could be user related (example: contact with sharp object)/ mechanical failure/ operator error/ thin rubberize layer.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "method for use: (1) do not inflate the balloon in the urethra.(the urethra may be injured).(2) do not pull the catheter hard.(the bladder/urethra maybe injured).2.Applicable patients: (1) patients with delirium who might pull out catheter (when patient tugs at catheter unconsciously, the bladder and urethra may be damaged).Contraindications: 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (4) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients: (1) do not use on patients who are or have been allergic to natural rubber latex." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRICATH FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10569742
MDR Text Key208528530
Report Number1018233-2020-06133
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0165L16
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/01/2020
Initial Date FDA Received09/22/2020
Supplement Dates Manufacturer Received12/24/2020
Supplement Dates FDA Received01/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-