ETHICON INC. BONE WAX; STERILE NONABSORBABLE BONE WAX Back to Search Results
Catalog Number W810T
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device w810t, lot lp7bjcm, and no non-conformances were identified.Investigation summary: it was reported packaging foreign matter non-biological.It was received for analysis a sealed sample of product code w810t, lot lp7bjcm.During the visual inspection of the sample, black foreign matter into the overwrap packet could be observed.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the sample condition the assignable cause is a foreign matter.The following information was requested, but unavailable: what was the initial procedure? not available.What is the event day and procedure date? not available.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please confirm: were there any adverse patient consequences? if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that a patient underwent an unknown procedure on an unknown and bone wax was used.It was noted that the product had black spots.There were no adverse patient consequences reported.Upon evaluation of the returned device, black foreign matter into the overwrap packet could be observed.Additional information has been requested.
 
Manufacturer Narrative
(b)(4).Date sent to the fda: 10/05/2020.Additional information was requested and the following was obtained: were there any adverse patient consequences? no patient consequence.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent to the fda: 09/28/2020 additional information: d4, h4 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE WAX
Type of Device
STERILE NONABSORBABLE BONE WAX
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key10574066
MDR Text Key208157002
Report Number2210968-2020-07232
Device Sequence Number1
Product Code MTJ
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 08/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Catalogue NumberW810T
Device Lot NumberLP7BJCM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2020
Date Manufacturer Received10/01/2020
Patient Sequence Number1
-
-