• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALSALVA AORTIC VALVED GRAFT; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALSALVA AORTIC VALVED GRAFT; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 25VAVGJ-515
Device Problems Fracture (1260); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020, a 25mm sjm masters series valsalva aortic valved graft was selected for implant in the patient.During the aortic root replacement procedure, the physician used the valve rotator recommended in the ifu, but one of the leaflets was damaged by rotating the discs.One of the leaflets disconnected from the hinge, the valve was removed from the patient, and a carbomedics valve was implanted instead.The patient remained stable throughout the procedure, but the procedure was extended by 30-45 minutes.The patient is currently stable.
 
Manufacturer Narrative
The reported event of a dislodged leaflet and damaged leaflet was confirmed.Two leaflets were received, one of which was fractured.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.While the root cause of the leaflet dislodgement could not be conclusively determined, there was no evidence of material defect in the carbon coating that may have caused or contributed to the dislodged leaflets or orifice damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALSALVA AORTIC VALVED GRAFT
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key10574969
MDR Text Key208199171
Report Number2648612-2020-00097
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734009492
UDI-Public05414734009492
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model Number25VAVGJ-515
Device Catalogue Number25VAVGJ-515
Device Lot Number7365362
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-