MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number RA002-5555SLR
Device Problem Defective Device (2588)
Patient Problem Failure of Implant (1924)
Event Date 02/15/2016
Event Type  Injury  
Manufacturer Narrative
The rod was received for visual and functional testing and the reported issue was confirmed as the device was not lengthening.Visual inspection revealed that the rod was positioning and was able to be lengthened by hand.Functional testing revealed that the locking pin was broken which resulted in the lead screw separating from the magnet cup module.
 
Event Description
Information was received that a revision procedure was performed on (b)(6) 2016.As per the reporter, the rod was removed for converting to final fusion and appeared to be jammed.The patient successfully completed treatment and lengthening sessions without issues; however, during the last distraction session, the rod did not appear to lengthen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
101 enterprise
suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
101 enterprise
suite 100
aliso viejo CA 92656
Manufacturer Contact
tiara rae
101 enterprise
suite 100
aliso viejo, CA 92656
6197314188
MDR Report Key10575402
MDR Text Key208192769
Report Number3006179046-2020-00398
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00856719002114
UDI-Public856719002114
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRA002-5555SLR
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age14 YR
Patient Weight38
-
-