Catalog Number 031-33J
Device Problem Gas/Air Leak (2946)
Patient Problem No Patient Involvement (2645)
Event Date 09/14/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
The complaint is reported as: "the user found air seemed to be leaking from the connecting part assembly to the flowmeter before use on a patient.Therefore, a new unit was used instead." no patient involvement.
 
Event Description
The complaint is reported as: "the user found air seemed to be leaking from the connecting part assembly to the flowmeter before use on a patient.Therefore, a new unit was used instead." no patient involvement.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was found that the puncture pin was missing.No other issues were found.In order to perform a functional inspection, the missing component (puncture pin) had to be assembled to the sample.The sample was then tested on the oxygen entrainment test with no functional issues.As an additional test the sample was tested on the oxygen entrainment test using a bottle of sterile water connected to the nebulizer adaptor and no air leaks were detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE
Type of Device
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key10580896
MDR Text Key208356459
Report Number3004365956-2020-00192
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
PMA/PMN Number
K153010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number031-33J
Device Lot Number74H1900578
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Date Manufacturer Received11/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
N/A.; N/A.
-
-