Model Number 37800 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Hypersensitivity/Allergic reaction (1907); Pain (1994); Swelling (2091); Vomiting (2144); Complaint, Ill-Defined (2331); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
Injury
|
Manufacturer Narrative
|
Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a health care provider( hcp) regarding a patient who was implanted with a neurostimulator for gastric stimulation.¿¿it was reported that during and emergency room visit, patient complained of pain at the ins site.Patient has an unrelated medical diagnosis of rumination syndrome.Impedance were checked and it was 650- within normal reading and voltage was 6.0.It was also stated that patient had inflammation, swelling, mucus and vomiting and hypersensitivity.The device was explanted and replaced.No further complications were reported/anticipated at this time.
|
|
Manufacturer Narrative
|
H3 :product analysis #(b)(6):analysis information -- 2021-(b)(6)14:38:39 cst pli# 10 product id# 37800 the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection, output and telemetry testing, and functional testing.Analysis determined that the implantable neurostimulator (ins) was functional.Continuation of d10: product id 4351-35 lot# serial# (b)(6) implanted: 2015-(b)(6) explanted: product type lead product id 4351-35 lot# serial# (b)(6) implanted: 2015-(b)(6) explanted: product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
No new information.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a healthcare professional (hcp).The hcp reported that the cause of the pain at the implant site was unknown.Removal of the system was performed (b)(6) 2018 without replacement.The patient did not benefit from the device and wanted to take the system out.Based on criteria, this patient did not qualify to receive the enterra at all; as the patient was diagnosed with "prumimetrium" (illegible) and not gastroparesis.
|
|
Event Description
|
Additional information was received from a healthcare professional (hcp).The hcp reported that the cause of the pain at the implant site was unknown.Removal of the system was performed on (b)(6) 2020 without replacement.The patient did not benefit from the device and wanted to take the system out.Based on criteria, this patient did not qualify to receive the enterra at all; as the patient was diagnosed with "prumimetrium" (illegible) and not gastroparesis.
|
|
Manufacturer Narrative
|
B5: please see correction of explant date as on (b)(6) 2020.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|