• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIANA CO. LTD. NELLCOR; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIANA CO. LTD. NELLCOR; OXIMETER Back to Search Results
Model Number 10005941J
Device Problem Defective Component (2292)
Patient Problems Death (1802); Insufficient Information (4580)
Event Date 09/04/2020
Event Type  Death  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, use of the device was terminated when the patient passed away.There was no allegation against the device in relation to the patient death.Subsequent routine servicing of the device replaced the battery, jog dial and the main body was upgraded.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the jog dial on the monitor needed to be replaced and the battery failed but was due to normal usage.The issue was resolved by replacing the jog dial and battery.It was reported that the jog dial on the pulse ox monitor needed to be replaced.The reported issue was confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the unit's jog dial was damaged.The units used was terminated because the patient passed away.There was no allegation against the monitor in relation to the patient death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Type of Device
OXIMETER
Manufacturer (Section D)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KR  00000
Manufacturer (Section G)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KR   00000
Manufacturer Contact
avi kluger
5920 longbow drive
boulder, CO 80301
3035306582
MDR Report Key10615113
MDR Text Key209441544
Report Number2936999-2020-00712
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120773
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10005941J
Device Catalogue Number10005941J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2020
Is the Reporter a Health Professional? No
Date Manufacturer Received11/03/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Death;
-
-