MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX TI 5.0 ULTRABRAID; STAPLE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX TI 5.0 ULTRABRAID; STAPLE, FIXATION, BONE Back to Search Results
Model Number 72200755
Device Problem Corroded (1131)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/29/2020
Event Type  malfunction  
Event Description
It was reported that during shoulder rotator cuff repair surgery, the inserter was found rust.It is unknown if there was a delay and a back up was available to complete the procedure.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Event Description
It was reported that during shoulder rotator cuff repair surgery, the inserter was found rust.No delay and a back up was available to complete the procedure.No other complications were reported.
 
Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The returned insertion device has reddish discoloration on the laser markings.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX TI 5.0 ULTRABRAID
Type of Device
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10618250
MDR Text Key209542524
Report Number1219602-2020-01508
Device Sequence Number1
Product Code JDR
UDI-Device Identifier03596010577986
UDI-Public03596010577986
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K972326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200755
Device Catalogue Number72200755
Device Lot Number2036676
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received12/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexMale
Patient Weight65 KG
-
-