Model Number 5866-38M |
Device Problems
Break (1069); Material Integrity Problem (2978)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2009 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: 4968-25 lead, implant date: (b)(6) 2004; a2dr01 ipg, implant date: (b)(6) 2016.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that while dissecting the pocket high undefined impedance and no capture at high output were noted on the right ventricular (rv) lead.Insulation degradation and tissue ingrowth were also noted on the y adapter.The adapter was explanted and the lead was partially removed.The implantable pulse generator (ipg) was explanted and replaced for an unknown reason.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that while dissecting the pocket high undefined impedance and no capture at high output were noted on the right ventricular (rv) lead.Insulation degradation and tissue ingrowth were also noted on the y adapter.The adapter was explanted and the lead was partially removed.The implantable pulse generator (ipg) was explanted and replaced for an unknown reason.No patient complications have been reported as a result of this event.It was also reported that the right atrial (ra) lead was fractured and that one rv lead was fractured and that both rv lead exhibited high thresholds.The leads remain in use.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was also reported that the right atrial (ra) lead was fractured and that one rv lead was fractured and that the other rv lead exhibited high thresholds.The leads remain in use.
|
|
Manufacturer Narrative
|
Product event summary: the full lead was returned, analyzed, and the outer insulation of the lead was extrinsically breached due to a cut.The proximal conductor of the lead became extrinsically distorted due to kinking/buckling.The distal conductor was extrinsically distorted due to kinking/buckling.Visual analysis of the lead indicated apparent explant damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was further reported that the device was explanted as it was nearing elective replacement indicator (eri).
|
|