GYRUS ACMI INC SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-T
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This device is an olympus demo asset returned for annual inspection.The device has been evaluated.The manufacture date is not known.Upon inspection and testing, the device had no functional problem.However, the device was observed to have a faulty connector and could not be locked on to the generator.
 
Event Description
This device was an asset return for annual inspection.There is no patient involvement.
 
Manufacturer Narrative
This report is updated in the following fields: d4, g4, g7, h2, h4, h6 and h10.The device had been previously serviced by olympus, therefore a service history review will replace device history record review.This device was recently estimated by olympus on february 25, 2020.At that time, cosmetic normal wear and tear was observed.The unit passed all functional and leakage current tests, all output powers were within specification.Damage to the connector is often incurred as a result of user error, often the user does not realize all connections are push/pull and instead the plug is twisted upon connection or removal.This action can result in damage to the pins internal to the socket.The device instructions for use states "connect the transducer to the generator by aligning the key way of the transducer connector with the key way slot on the transducer receptacle on the front panel.Push straight in.Caution do not twist or turn the plug.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI INC
136 turnpike road
southborough PA 01772
MDR Report Key10619996
MDR Text Key209594514
Report Number3011050570-2020-00092
Device Sequence Number1
Product Code FEO
Combination Product (y/n)N
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2020
Was the Report Sent to FDA? No
Date Manufacturer Received01/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-