• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C8320
Device Problem Filling Problem (1233)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/07/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a home patient (hp) received an over delivery of solution during therapy with a homechoice pro automated pd system.This was described as the pd registered nurse ¿gave five liters of infusion to the hp and there were no alarms that occurred¿.The hp was connected at the time of the event.This occurred during an unspecified process step of peritoneal dialysis (pd) therapy.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information in d10, h3, h6.H10: the device was received for evaluation.A simulated therapy was successfully performed on the device.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10620101
MDR Text Key209589968
Report Number1416980-2020-06145
Device Sequence Number1
Product Code FKX
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberR5C8320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2020
Date Manufacturer Received10/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
HOMECHOICE CASSETTE; UNSPECIFIED SOLUTION; HOMECHOICE CASSETTE; UNSPECIFIED SOLUTION
-
-