• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ENSITE VELOCITY CARDIAC MAPPING SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ENSITE VELOCITY CARDIAC MAPPING SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number EE3000
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/01/2020
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
During preparation with the patient on the table, the amplifier was booted up, displayed a flashing amber light and the case was cancelled.Multiple reboots were attempted, the air filter was cleaned, and a second outlet was attempted.The issue was not able to be resolved and the procedure was aborted and cancelled.
 
Manufacturer Narrative
One ensite velocity¿ system velocity amplifier was received for analysis.Visual inspection of the returned product confirmed all input and output connectors are free of physical damage and all labeling is legible and correctly oriented.Ac power was then applied, and the returned amplifier failed to successfully complete the post (power on self-test) and the system status light remained amber.Additional analysis isolated the root cause of the reported issue to an incomplete system calibration of the catheter amplifier pca (printed circuit assembly) located in the slot 11 position.Failure to power up slot 11 was also confirmed within the system log files.The system was recalibrated, and normal functionality was restored to the system.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information provided to abbott and the investigation performed, the root cause of the reported event has been isolated to an incomplete system calibration which resulted in the loss of the catheter amplifier board at the slot eleven location.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY CARDIAC MAPPING SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key10622938
MDR Text Key209775337
Report Number2184149-2020-00166
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05414734210713
UDI-Public05414734210713
Combination Product (y/n)N
PMA/PMN Number
K160187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberEE3000
Device Catalogue NumberEE3000
Device Lot Number7052735
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-