• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTH,HIP,SEMCONSTR,UNCEMN,MET/POLY,NONPOR,CALPHOSP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTH,HIP,SEMCONSTR,UNCEMN,MET/POLY,NONPOR,CALPHOSP Back to Search Results
Model Number 71360093
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 09/08/2020
Event Type  malfunction  
Event Description
It was reported that the tip of the femoral head/neck impactor was found to be broken.No case related therefore no patient involved.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure.The device tip is missing.The device was manufactured in 2018 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD/NECK IMPACTOR
Type of Device
PROSTH,HIP,SEMCONSTR,UNCEMN,MET/POLY,NONPOR,CALPHOSP
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10623412
MDR Text Key209750294
Report Number1020279-2020-05110
Device Sequence Number1
Product Code MEH
UDI-Device Identifier03596010619310
UDI-Public03596010619310
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360093
Device Catalogue Number71360093
Device Lot Number18CM14233
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Date Manufacturer Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-