Investigation results were made available.Event description: it was reported that during surgery, the durasul alpha insert could not be inserted into the cup.Despite several attempts, the insert could not be inserted into the cup to make it hold.Surgery was completed with another new inlay.Product evaluation: visual examination: the complained durasul alpha liner was returned for investigation.Articulation side: the rim shows some scratches and imprints from the liner setting instrument.Further, on the articulation surface some scratches and nicks can be seen.Backside: the centralizing liner pole pin is cut off.Signs of attempts of fitting the insert in the shell can be seen in the form of indented lines and scratches across the anchoring side.Based on the visual examination the reported event can be confirmed.Measurements: to ensure the durasul alpha inlay hooded kk/32 have correct dimensions, relevant characteristics according to inspection plan were measured with caliper rüegger.Characteristic no.3 feature (diameter 50.37 mm +/- 0.05).Specification: max.50.42 mm; min.50.32 mm.Measured value: 50.39 +/- 0.02 mm.Conclusion: the characteristic is within specification.Review of product documentation: device purpose: this device is intended for treatment.Dhr review: review of the device history records identified no deviations or anomalies during manufacturing.Ncr(s): no ncr with a potential correlation to the reported event was found.Surgical technique: (if the liner can still be rotated after light impaction, this indicates mispositionning, nonconcentric, or soft tissues interference between the liner and the shell surfaces.) (if the fitting of the insert is faulty, a new insert must be used.If the polar peg is deformed, it will not be possible to anchor the insert correctly.) (note: if unsure, the seating can be checked using a raspatory.If the fitting of the liner is faulty, a new insert must be used.If the polar peg is deformed, it will not be possible to anchor the liner correctly.) conclusion: it was reported that during surgery, the durasul alpha insert could not be inserted into the cup.Despite several attempts, the insert could not be inserted into the cup to make it hold.Surgery was completed with another new inlay.Based on the investigation the reported event can be confirmed.The detected damages / imperfections on the returned insert indicate that the liner was mispositioned, nonconcentric or interfered with soft tissue residues.It is possible that remains of tissue and/or inappropriate alignment of the inlays prior or during impaction prevented the inlays from snapping into position.Further, cutting off the pole pin of the inlays is considered an off-label use, as the durasul alpha inlay must not be altered in any way.The pole pin is an essential design feature for centralizing and stabilizing the inlay inside the allofit shell.Nevertheless, based on the investigation and the available information, we were not able to identify an exact root cause for the reported assembly issue.However, this is a common complication also described in the applicable surgical technique of the allofit acetabular system.This complication requires the use of a new liner, which has been applied correctly.The investigation did not identify a nonconformance or a complaint out of box (coob).The need for corrective measures is not indicated and zimmer switzerland manufacturing gmbh considers this case as closed.Zimmer biomet's reference number of this file is (b)(4).
|