ZIMMER BIOMET, INC. AUGMENT REAMER DRIVER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number SBGL3701
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/09/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported while using the augment reamer driver for the first time, the soccer ball tip fractured off.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D4: (b)(4).Visual examination of the returned product identified the item fractured.Hardness taken is within specification.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
Upon reassessment of the reported event, it was determined to be not reportable.The initial reports were forwarded in error and should be voided.
 
Manufacturer Narrative
Upon reassessment of the reported event, it was determined to be not reportable.The initial reports were forwarded in error and should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUGMENT REAMER DRIVER
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10635642
MDR Text Key210013812
Report Number0001822565-2020-03383
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
PMA/PMN Number
K193180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup,Followup
Report Date 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSBGL3701
Device Lot Number64720886
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-