Catalog Number SBGL3701 |
Device Problem
Fracture (1260)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported while using the augment reamer driver for the first time, the soccer ball tip fractured off.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No further event information is available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.D4: (b)(4).Visual examination of the returned product identified the item fractured.Hardness taken is within specification.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
Upon reassessment of the reported event, it was determined to be not reportable.The initial reports were forwarded in error and should be voided.
|
|
Manufacturer Narrative
|
Upon reassessment of the reported event, it was determined to be not reportable.The initial reports were forwarded in error and should be voided.
|
|