Model Number 71992-01 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Sweating (2444); Palpitations (2467); Shaking/Tremors (2515)
|
Event Date 09/27/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kit were reviewed and the dhrs showed the libre sensor and libre sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A caller reported a customer received an error message, "scan again in 10 minutes", for 2 hours or more with the adc freestyle libre 2 sensor.On (b)(6) 2020, the customer experienced sweating, trembling, and palpitations and was unable to treat themselves.The customer was provided unspecified treatment by his wife for his symptoms.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug was properly seated and no issues were observed with the returned sensor patch.Extracted data from the returned sensor using approved software.Sensor found to be in sensor state 5 (indicating normal termination).Visually inspected the plug assembly and no issues were observed.A sim vivo test was performed and the results were with in specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.Additional information- section g1: (contact office first name, contact office last name, contact office phone number and contact office email) have been updated to (b)(6).
|
|
Event Description
|
A caller reported a customer received an error message, "scan again in 10 minutes", for 2 hours or more with the adc freestyle libre 2 sensor.On (b)(6)2020, the customer experienced sweating, trembling, and palpitations and was unable to treat themselves.The customer was provided unspecified treatment by his wife for his symptoms.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|