The device, used in treatment, was returned for evaluation.The cutting block exhibited wear in each of the cutting slots and damage to the impactor connection point which are consistent with product use.The cutting block bone spikes fractured at the cutting block interface.The fracture of the bone spikes is likely due to stresses in excess of its material properties being applied to the bone spike during use.The broken spikes were not returned.The device was manufactured in 2009.A medical investigation was conducted and this case reports that when removing the femoral cutting block, the securing pins fractured and remained in the femur.Per email communication, the pieces were removed with a forceps with no injury to the patient, and the procedure was completed with a delay of 15 minutes.Since no patient harm is alleged, no further clinical assessment is warranted at this time.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
|