• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO SECURE MRI SURESCAN; DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO SECURE MRI SURESCAN; DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 6947M62
Device Problems Positioning Failure (1158); Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/28/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the implant procedure the right ventricular (rv) lead helix would not extend or retract.Another lead was used to complete procedure.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.The analysis indicated that the helix exceeded specification the number of turns to extend and retract.The analyst noted the full lead was returned with a stylet in the lead.The helix does not extend due to the driveshaft lug being stuck on the retraction stop.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE MRI SURESCAN
Type of Device
DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key10656009
MDR Text Key210623632
Report Number2649622-2020-19555
Device Sequence Number1
Product Code NVY
UDI-Device Identifier00643169356627
UDI-Public00643169356627
Combination Product (y/n)N
PMA/PMN Number
P920015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/15/2021
Device Model Number6947M62
Device Catalogue Number6947M62
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Date Manufacturer Received10/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age65 YR
-
-