Catalog Number A1059 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A customer reported that the lock of the a1059 mayfield modified skull clamp was hard to turn.There was no known surgery delay or patient injury.
|
|
Manufacturer Narrative
|
Unique device identifier (udi): (b)(4).The mayfield skull clamp was returned for evaluation: device history record (dhr) - no abnormalities related to the reported failure shown in dhrs found for lot# 124.Complaint confirmed via inspection of the unit.Unit received with the lock having rotational and lateral movement and was difficult to lock due to wear.Unit needed new components added to replace worn internal parts.With respect to the complaint, it was confirmed the returned unit did not meet all specific functional tests.
|
|
Event Description
|
N/a.
|
|