MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFAPRO28
Device Problems Gas/Air Leak (2946); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem ST Segment Elevation (2059)
Event Date 09/24/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, st elevation occurred.Right ventricular pacing was performed and medications were administered to mitigate the st elevation.The physician believed air ingress occurred during the procedure.The st elevation recovered and the case was completed with cryo.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed at least nine applications were performed with a non-returned afapro28 balloon catheter with lot 56454 without any issue on the date of the event.In conclusion, the clinical issue (st elevation) occurred during procedure.There is no indication of relation of adverse event to the performance of the product.The air ingress issue cannot be assessed through data analysis.The physical product is not available for further analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE PRO CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key10661577
MDR Text Key211301451
Report Number3002648230-2020-00544
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/10/2021
Device Model NumberAFAPRO28
Device Catalogue NumberAFAPRO28
Device Lot Number56454
Was Device Available for Evaluation? No
Date Manufacturer Received10/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
4FC12 SHEATH
Patient Outcome(s) Required Intervention;
-
-