Catalog Number C2602
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/11/2020
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A customer reported that the physician assembled the c2602 cusa excel 36khz straight handpiece and when he torqued the click, it did not sound.The unit did not pass the amplitude test.There was no patient injury and no known surgery delay.Additional information has been requested.
 
Event Description
N/a.
 
Manufacturer Narrative
Unique device identifier (udi): (b)(4).The cusa handpiece was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSA EXCEL 36KHZ STRAIGHT HANDPIECE
Type of Device
ULTRASONIC SURGICAL PRODUCTS
Manufacturer (Section D)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
ida business&technology park
sragh, tullamore, co.offaly
EI 
MDR Report Key10665988
MDR Text Key214801096
Report Number3006697299-2020-00112
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
PMA/PMN Number
K141668
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC2602
Was Device Available for Evaluation? No
Date Manufacturer Received10/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-