The device, used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is cracked and damaged and has signs of wear and tear from use.As per clinical/medical investigation, trial cracked in half during the procedure.It was communicated that the navio procedure was completed as planned without a retained foreign body, surgical delay, or additional medical interventions.Reportedly, the patient status is fine.Based on the information provided, further patient impact would not be anticipated as the procedure was completed as planned without delay or patient harm.No further medical assessment is warranted at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.
|