• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND MICRA; LEADLESS PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND MICRA; LEADLESS PACEMAKER Back to Search Results
Model Number MC1VR01
Device Problem Battery Problem (2885)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a leadless implantable pulse generator (ipg) could not be implanted as estimated remaining longevity was not available, possibly related to storage of device at low temperature.The device was not implanted.There was no patient involvement in this event.
 
Event Description
It was further reported that a new leadless ipg was implanted.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.Hybrid analysis found the cause of the low battery voltage was not determined.Electrical testing of the hybrid determined it was fully functional with no hybrid anomalies identified.The cause of the low battery voltage was not determined.Battery analysis during destructive analysis found no evidence of a separator breach but discoloration at the ethylene tetrafluoroethylene/glass interface of the feedthrough was observed.Glass reflow on the interior side of the feedthrough was confirmed by the scanning electron microscope examination, which indicted an abnormally high energy effect from the battery electrical connector b+ laser weld.This type of thermal damage has been known to induce an internal shorting path and was the likely cause of the ¿low battery voltage out of the box¿.The anomalies observed on the feedthrough may have contributed to depletion mechanisms observed on similar returns per capa 439503.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRA
Type of Device
LEADLESS PACEMAKER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key10674227
MDR Text Key219375075
Report Number9612164-2020-03893
Device Sequence Number1
Product Code PNJ
Combination Product (y/n)N
PMA/PMN Number
P150033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2021
Device Model NumberMC1VR01
Device Catalogue NumberMC1VR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Date Manufacturer Received09/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-