Catalog Number 00MLX |
Device Problem
Excessive Heating (4030)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A customer reported that the 00mlx mlx 300w xenon lightsource has a fan issue and overheats.There was no known patient contact or surgery delay.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Unique device identifier (udi): (b)(4).Dhr review: dhr shows no abnormalities related to the reported issue.The returned 00mlx light source was in used condition with a disconnected connection on the fan preventing proper function.The reported complaint is confirmed.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|