Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
Device was returned with packaging damage, which compromised sterility barrier.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual evaluation of the returned product found the carton is crushed near the perforation, and the perforation is opened.Both inner and outer cavities are cracked and crushed.The sterility has been compromised.The complaint is confirmed.Device history record (dhr) was reviewed and no discrepancies were found.The root cause of the reported issue is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0 MM DIAMETER CORTICAL SCREW - RED FIXED ANGLE 3.5 MM HEX HEAD
Type of Device
PROSTHESIS, TRAUMA
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10677539
MDR Text Key211368845
Report Number0001822565-2020-03530
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
PMA/PMN Number
K082770
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Model NumberN/A
Device Catalogue Number47248410050
Device Lot Number61803252
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-