Model Number 8900-4005
Device Problem Arcing of Electrodes (2289)
Patient Problem No Information (3190)
Event Date 10/01/2020
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation, and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), a spark was seen from the electrode pads while attached to the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The electrodes were returned to zoll medical corporation; evaluation of the electrodes revealed that there was some hair and skin particles stuck to the pads.It is suspected that the cause of the reported event is due to poor coupling of the electrode pads to the patient.The ifu states poor adherence and/or air under the electrode can lead to possibility of arcing and skin burns.Good skin preparation does not guarantee a burn will not occur and burns from defibrillation is an expected risk.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO PADZ, MFE, RADIOLUCENT, ADULT, HVP
Type of Device
DEFIB/PACING ELECTRODE
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
MDR Report Key10680211
MDR Text Key211365116
Report Number1218058-2020-00080
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
PMA/PMN Number
K960676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/18/2021
Device Model Number8900-4005
Device Catalogue Number8900-4005
Device Lot Number2520B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-