ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Death (1802); Bone Fracture(s) (1870); Burn, Thermal (2530); Shock from Patient Lead(s) (3162)
Event Date 08/30/2020
Event Type  Death  
Manufacturer Narrative
Device evaluation summary: monitor sn (b)(4) and electrode belt sn (b)(4) have been returned to zoll manufacturing corporation, but have not been evaluated yet.Device evaluation included review of downloaded software flag files on the day of the event.The review of the software flags consisted of an analysis of the downloaded data to identify any fault flags or unusual patterns of software flags.The software flag files did not suggest a device malfunction that would contribute to the inappropriate treatment.Treatment event analysis concludes: the patient received five inappropriate shocks.Amplitude oversensing and multiple counting contributed to the false detections.The response buttons were not pressed during the event.Patient's death was reported on (b)(6) 2020.Additional inappropriate defibrillation narrative: inappropriate defibrillations are an anticipated risk associated with the use of the lifevest.Patients are instructed through alarms, voice messages, ifu, and training to press the response buttons to prevent an inappropriate defibrillation.The current commercial inappropriate defibrillation rate is consistent with the observed rate during the pivotal clinical trial (b)(4) (0.69% per patient-month with 90% confidence).A summary of the safety and effectiveness data (ssed), including the inappropriate defibrillation safety objective supporting fda's approval of the lifevest, can be found at (http://www.Accessdata.Fda.Gov/cdrh_docs/pdf/p010030b.Pdf).The lifevest detection algorithm complies with iec 60601-2-4 performance requirements for sensitivity and specificity.
 
Event Description
A us distributor contacted zoll to report that a (b)(6) patient passed away while wearing the lifevest on (b)(6) 2020.The patient was sedated in the hospital at the time of passing.Review of the download data indicates the patient received five inappropriate shocks on (b)(6) 2020 in response to amplitude oversensing and multiple counting.The patient was in sinus rhythm at 80-90 bpm at the time of each of the five shocks at 20:18:54, 20:27:37, 20:39:41, 20:41:30, and 20:45:19.The post shock rhythm for each shock was sinus rhythm at 80-90 bpm.The response buttons were not pressed during the treatment event.It was reported that the patient was terminally ill and rejected any life-prolonging measurements.After the patient received the treatments, it was reported that the treatments caused the patient to get fractured ribs and burned skin.There is no indication the alleged injury lead to a serious deterioration in the health of the patient, as the patient was reported to be terminally ill prior to the treatments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key10686455
MDR Text Key211562926
Report Number3008642652-2020-08954
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0988
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2020
Date Manufacturer Received09/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death; Hospitalization; Other;
Patient Age74 YR
-
-