Model Number CI-1601-04 |
Device Problems
Break (1069); Disconnection (1171); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 10/22/2020 |
Event Type
Injury
|
Event Description
|
The recipient is reportedly experiencing sound quality issues due to a broken electrode array.According to the recipient's surgeon, a ct scan confirmed discontinuity at the proximal end of the electrode array.Programming adjustments were made, however, the issue did not resolve.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was reimplanted with another cochlear implant.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|