MAUDE Adverse Event Report: SMITH & NEPHEW, INC. STRAIGHT SCREWDRIVER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. STRAIGHT SCREWDRIVER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71362293
Device Problems Defective Device (2588); Naturally Worn (2988)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2020
Event Type  malfunction  
Event Description
It was reported that during efip inspection the straight screwdriver was found with a worn out end.No case related therefore no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation a visual inspection confirms the straight screwdriver is stripped and rounded off on both ends of the device.The device shows extensive signs of wear/usage.The device was manufactured in 2017.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT SCREWDRIVER
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10701802
MDR Text Key212048262
Report Number1020279-2020-05536
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010454188
UDI-Public03596010454188
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71362293
Device Catalogue Number71362293
Device Lot Number17DM15095
Was Device Available for Evaluation? No
Date Manufacturer Received11/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-