Catalog Number 397003-001 |
Device Problem
Failure to Calibrate (2440)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion hospital cart will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The customer, a syncardia certified hospital, reported that while supporting a patient the companion hospital cart's touch screen did not calibrate.There was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
The companion hospital cart passed all functional testing.Additionally, the hospital cart was tested using two known functioning companion 2 drivers in an attempt to confirm and replicate the customer-reported issue that the touch screen did not calibrate.The hospital cart passed all calibration tests.The root cause of the customer-reported issue could not be determined.The companion hospital cart performed as intended with no evidence of a device malfunction.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.Ce 5376 follow-up report 1.
|
|