ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA) EPIC SUPRA VALVE W/FLEXFIT; HEART-VALVE, NON-ALLOGRAFT TISSUE
|
Back to Search Results |
|
Model Number ESP100-25 |
Device Problem
Incomplete Coaptation (2507)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/02/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Further information regarding this event has been requested.Investigation results will be provided in a subsequent submission.
|
|
Event Description
|
On (b)(6) 2020, a epic valve was chosen form procedure.After implanting the epic supra valve (esp100-25), it was noticed in the echo, that the valve didnt close properly.The valve was then explanted and a new valve was implanted.No patient consequences were reported.
|
|
Manufacturer Narrative
|
The reported event the valve not closing properly could not be confirmed.No anomalies were found with the valve cusps, and functional testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper cuspal coaptation and hemodynamic performance.Please note, the free state of the leaflets, without physiological pressure applied, is not indicative of actual leaflet coaptation or valve function.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
|
|
Event Description
|
On (b)(6) 2020, a epic valve was chosen form procedure.During the implantation of the initial device, after implanting the epic supra valve (esp100-25), it was noticed by the physician that the valve didn't close properly.Initial inspection of the device showed a slight retraction of one cusp.The valve was exchanged during the same operation procedure, the valve was then explanted and a new valve was implanted.No patient consequences were reported.
|
|