• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA) EPIC SUPRA VALVE W/FLEXFIT; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA) EPIC SUPRA VALVE W/FLEXFIT; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number ESP100-25
Device Problem Incomplete Coaptation (2507)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.Investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020, a epic valve was chosen form procedure.After implanting the epic supra valve (esp100-25), it was noticed in the echo, that the valve didnt close properly.The valve was then explanted and a new valve was implanted.No patient consequences were reported.
 
Manufacturer Narrative
The reported event the valve not closing properly could not be confirmed.No anomalies were found with the valve cusps, and functional testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper cuspal coaptation and hemodynamic performance.Please note, the free state of the leaflets, without physiological pressure applied, is not indicative of actual leaflet coaptation or valve function.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Event Description
On (b)(6) 2020, a epic valve was chosen form procedure.During the implantation of the initial device, after implanting the epic supra valve (esp100-25), it was noticed by the physician that the valve didn't close properly.Initial inspection of the device showed a slight retraction of one cusp.The valve was exchanged during the same operation procedure, the valve was then explanted and a new valve was implanted.No patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC SUPRA VALVE W/FLEXFIT
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA)
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR  31310-260
MDR Report Key10706144
MDR Text Key212134619
Report Number3001743903-2020-00011
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/11/2023
Device Model NumberESP100-25
Device Lot NumberBR00023220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
-
-