Model Number ER420
Device Problems Failure to Form Staple (2579); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/16/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # u9493u.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during an unknown procedure, the clip was not fed into the jaws properly during the second firing, and the deployed clip was not formed properly.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Manufacturer Narrative
(b)(4).Date sent: 11/10/2020 d4: batch # u9493u investigation summary the analysis results found that the er420 device was returned with no apparent damage.The device was tested for functionality.During the analysis, the device was cycled, and it fed, retained, and formed the remaining one clip as intended.The event described could not be confirmed as the device performed without any difficulties noted.Although there is no direct evidence of use error, the instructions for use do contain the following: "caution: do not excessively twist or torque the instrument jaws when positioning the instrument on a vessel and firing.Excessive twisting or torquing may result in clip malformation." it should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device lot/batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key10711475
MDR Text Key212297050
Report Number3005075853-2020-05434
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012597
UDI-Public10705036012597
Combination Product (y/n)N
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberER420
Device Catalogue NumberER420
Device Lot NumberU9493U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Date Manufacturer Received10/21/2020
Patient Sequence Number1
-
-