Model Number EL5ML
Device Problem Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # u9409e.Investigation summary: the analysis results found that the el5ml device was returned with no apparent damage.In addition, two clips malformed were received inside of a plastic bag.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 13 conforming clips.The event described could not be confirmed as the device performed without any difficulties noted.It should be noted that as part of our quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's were identified.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during an unknown procedure, there was an issue with the el5ml firing two clips at a time.There were no patient consequences.No additional information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key10714792
MDR Text Key212491964
Report Number3005075853-2020-05454
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberU9409E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/24/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-