(b)(4).Batch # u9409e.Investigation summary: the analysis results found that the el5ml device was returned with no apparent damage.In addition, two clips malformed were received inside of a plastic bag.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 13 conforming clips.The event described could not be confirmed as the device performed without any difficulties noted.It should be noted that as part of our quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's were identified.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|