• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. T7 LINEAR DRIVER SHAFT W/ AO QC; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. T7 LINEAR DRIVER SHAFT W/ AO QC; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE Back to Search Results
Model Number 71174985
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/24/2020
Event Type  malfunction  
Event Description
It was reported during the procedure that the tip of device was twisted.The procedure was completed without delay and backup from smith& nephew was available to finish the surgery.No patient injury or other complications were reported.
 
Manufacturer Narrative
Results of investigation: the device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T7 LINEAR DRIVER SHAFT W/ AO QC
Type of Device
INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10715034
MDR Text Key212443091
Report Number1020279-2020-05632
Device Sequence Number1
Product Code MDM
UDI-Device Identifier00885556500545
UDI-Public00885556500545
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71174985
Device Catalogue Number71174985
Date Manufacturer Received03/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-