Bioburden released from inside of dilator from lite decompression tube tray when 18mm tube was passed through dilator during posterior cervical minimally invasive partial laminectomy and foraminotomy at left c6-c7.Patient received 3gm ancef intraoperatively.Found particles on seals and dilators.Checked the parts for bio-burden.After cleaning the parts, swabbed inside and found brown/grey residue.Expanded testing to other devices and also determined stryker tubes had a similar residue inside.To test inside, used a sterile cotton swab and sterile water to scrub.Cotton tip looked brown/grey.Tested all tubes and dilators with a dry sterile swab, then a moist (sterile water) sterile swab, then a moist (70% alcohol) sterile swab.Each time, residue was found.Provided a new tube set from inventory.Swab test confirmed a residue in the new set.The new set has since been cleaned and is being used at the hospital.A high sense of urgency is needed by stryker to support this investigation.Stryker: we are committed to urgently support this investigation and get good parts to the hospital for usage.
|