Model Number HL-390 |
Device Problems
Contamination (1120); Fluid/Blood Leak (1250)
|
Patient Problems
No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
One level 1 low flow fluid warming system was received with a cracked enclosure at the drain fitting, cracked tanks, the heater did not pass electrical testing and the line cord was worn.The tank was filled with water, a temperature check was attached, the line cord was plugged and the device was turned on.The reported leaking issue was unable to be duplicated.After running for several hours the device had no signs of residue in the reservoir and did not leak anywhere.The faulty heater was replaced.There was no fault found with the returned device.
|
|
Event Description
|
Information was received indicating that a smiths medical level 1 low flow fluid warming system was leaking and had residue inside the container.There was no reported adverse event.
|
|
Manufacturer Narrative
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
|