• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Device Problem Material Puncture/Hole (1504)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/01/2020
Event Type  malfunction  
Event Description
It was reported that balloon shaft hole occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was selected for use.However, it was noted that the balloon had a small hole at the base of the shaft.The procedure was completed with another of the same device.No patient complications reported and the patient's status was fine.
 
Manufacturer Narrative
Device evaluated by mfr.: gladiator 10 x 40, 75cm was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 12mm proximal from the proximal markerband.The rated burst pressure for this device is 14 atmospheres.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.No other issues were identified during the product analysis.
 
Event Description
It was reported that balloon shaft hole occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was selected for use.However, it was noted that the balloon had a small hole at the base of the shaft.The procedure was completed with another of same device.No patient complications reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10722203
MDR Text Key212987712
Report Number2134265-2020-14163
Device Sequence Number1
Product Code LIT
UDI-Device Identifier08714729809517
UDI-Public08714729809517
Combination Product (y/n)N
PMA/PMN Number
K132810
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/15/2023
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0025516324
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2020
Date Manufacturer Received10/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-