• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100H-12
Device Problems Loose or Intermittent Connection (1371); Mechanical Problem (1384); Intermittent Communication Failure (4038)
Patient Problems Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/21/2021
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing loss of lock.Device testing was attempted, however, lock could not be achieved.Revision surgery is being considered.
 
Manufacturer Narrative
Additional information: section b.6.External equipment was exchanged, however, the issue did not resolve.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Additional information: sections b.3, d.6b.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode array was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.It is believed that the failure of this implantable cochlear stimulator (ics) device was caused by a loss of hermetic seal, which was concluded from the residual gas analysis test data.This ultimately caused the device to cease functioning.This older device configuration is no longer manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
alexandra gonzalez
28515 westinghouse place
valencia, CA 91355
MDR Report Key10726516
MDR Text Key214535597
Report Number3006556115-2020-01168
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeES
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/01/2003
Device Model NumberAB-5100H-12
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2021
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2001
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age36 YR
Patient SexFemale
-
-